Abstract

An HTS SQUID gradiometer consisting of a series gradiometer pickup loop in flip-chip configuration with a small washer SQUID is presented. The series gradiometer configuration is advantageous for applications without shielding because of the avoidance of circulating shielding currents in the pickup loop. With a 20 mm × 10 mm substrate for the pickup loop a large effective area of 1.25 mm2 and a gradiometer baseline of 9.6 mm are achieved. The balance of the gradiometer impaired by the washer read-out SQUID can significantly be enhanced by the weighted subtraction of a reference SQUID's signal. With this a balance of 5 × 104 is reached. The noise-limited magnetic field gradient resolution of the balanced SQUID gradiometer is 38 fT cm−1 Hz−1/2.This paper was presented at the 8th International Superconductive Electronics Conference, Osaka, Japan, 19–22 June 2001.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.