Abstract

Introduction1,1-Difluoroethane (HFA-152a) is being developed as an alternative propellant in pressurized metered dose inhalers (pMDIs). As a part of the regulatory development pathway, pharmacology, toxicology and clinical studies have been conducted with inhaled HFA-152a. These studies require fit for purpose regulatory compliant (GxP validated) methods for quantification of HFA-152a from blood. MethodsAs HFA-152a is a gas at standard temperature and pressure, novel methods were developed to support the analysis across the wide range of species and concentrations required for regulatory filing. ResultsThe developed methods utilized a headspace auto sampler coupled to a gas chromatograph (GC) with flame ionization detection. Key factors in the successful method included bringing together fit for purpose approaches to the head space vials, volume of matrix (blood), detection range required for species/study objective, handling / transfer of blood into head space vials and the stability/storage required for the analysis of the samples. The species-specific assays were fully validated under regulatory (GLP) conditions for mouse, rat, rabbit, canine and human and non-regulatory (non GLP) validations for guinea pig and cell culture media. DiscussionOverall the novel approach of head space analysis of whole blood allowed for the development and validation of assays used to generate the toxicokinetic data that supported clinical testing of HFA-152a as a new pMDI propellant.