Abstract

The High-Resolution Energetic X-ray Imager (HREXI) cadmium zinc telluride (CZT) detector development program at Harvard is aimed at developing tiled arrays of finely pixelated CZT detectors for use in wide-field coded aperture 3 to 200 keV x-ray telescopes. A pixel size of ∼600 μm has already been achieved in the ProtoEXIST2 (P2) detector plane with CZT readout by the NuSTAR application-specific integrated circuits. This paves the way for even smaller than 300-μm pixels in the next-generation HREXI detectors. We describe a new HREXI calibration facility (HCF) that enables a high-resolution subpixel-level (100 μm) two-dimensional (2D) scan of a 256-cm2 tiled array of 2 × 2 cm2 CZT detectors illuminated by a bright x-ray AmpTek Mini-X tube source at timescales of around a day. HCF is a significant improvement from the previous apparatus used for scanning these detectors, which took ∼3 weeks to complete a one-dimensional (1D) scan of a similar detector plane. Moreover, HCF has the capability to scan a large tiled array of CZT detectors (32 × 32 cm2) at 100-μm resolution in the 10- to 50-keV energy range, which was not possible previously. We describe the design, construction, and implementation of HCF for the calibration of the P2 detector plane.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.