Abstract

Abstract A precise and accurate HPLC method for the quantification of ketorolac in human plasma is described. Ketorolac and an added internal standard (a fluoro analog of ketorolac) are extracted from acidified plasma samples and the extract is further purified. The reconstituted sample is injected onto a reverse phase HPLC column. The method has a linear concentration range of 0.010-3.00 μg of ketorolac per aliquot of plasma, using 0.1 to 1.0 mL of plasma for analysis. The quantification limit of the method is 0.010 μg/mL. The intra- and inter-assay % CVs were less than 5.5%, and the mean recoveries ranged from 96% to 109%. The absolute recovery was 92%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.