Abstract

A fast and sensitive high performance liquid chromatography method for quantitative determination of rosiglitazone in human plasma has been developed. The extraction from plasma was performed using solid-phase extraction (SPE) on C4 silica (100 mg) disposable extraction cartridges (DEC). The separation of rosiglitazone and two metabolites was achieved on a Phenomenex® Synergi 4 µm MAX-RP (150 × 4.6 mm) column, protected by a guard column. The mobile phase was 0.01 M ammonium acetate, pH 7.0 - acetonitrile (65:35, v/v). (3S)-3-OH-quinidine was used as internal standard. The analytes were detected using fluorescence detection. The method was validated. The limit of quantitation was 1 ng mL−1 and the detection limit was 0.25 ng mL−1 for rosiglitazone in human plasma. The recovery was 90% for rosiglitazone. Linearity was observed over a range of 1-1000 ng mL−1 (r2=0.9959). The intra- and inter-day precision (C.V.) did not exceed 8.7 %. Applicability of the method was demonstrated by a clinical pharmacokinetic study. A healthy volunteer received in two separate phases 4 mg and 8 mg rosiglitazone maleate as a single oral dose. Plasma concentrations were measured for 24 h in both phases.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.