Abstract

Introduction: Studies have shown when hemodialysis treatment is sufficiently effective, complications of uremic syndrome, additional treatment costs and hospitalization length are reduced. Several methods have been proposed to improve hemodialysis adequacy. Objectives: In this study, the effects of the synchronic use of the stepwise profile dialysis fluid flow rate with increased blood flow rate (BFR) were studied on hemodialysis adequacy. Patients and Methods: This is a cross-over clinical trial study conducted on 34 hemodialysis patients selected from a hemodialysis center of Qazvin University of Medical Sciences, Qazvin, Iran. The patients were randomly allocated into two groups (n= 17 patients in each group) in two sequences. In the first sequence, the subjects received four routine hemodialysis sessions in group one and four hemodialysis sessions with the stepwise profile of the fluid flow rate with increased BFR in group two. In the second sequence, the treatment methods were exchanged. Hemodialysis sessions were performed in both sequences, consecutively. Each session was at least three hours. Hemodialysis adequacy was measured using Kt/V software on the hemodialysis machines after each session. Results: The mean score of dialysis adequacy was 0.89 in the routine method and 1.26 in the profile with increased BFR. There was a statistically significant difference between the methods (t= -7.9, df = 33, P < 0.001). Conclusion: The results of the study suggest that the stepwise profile of the dialysis fluid flow rate with increased BFR should be used synchronously to improve hemodialysis adequacy.

Highlights

  • Studies have shown when hemodialysis treatment is sufficiently effective, complications of uremic syndrome, additional treatment costs and hospitalization length are reduced

  • The results of this study showed that the synchronic use of the stepwise profile dialysis fluid flow rate with increased blood flow rate improved hemodialysis adequacy

  • In the second group (17 patients), a significant difference was reported in terms of hemodialysis adequacy between the www.nephropathol.com

Read more

Summary

Introduction

Studies have shown when hemodialysis treatment is sufficiently effective, complications of uremic syndrome, additional treatment costs and hospitalization length are reduced. Objectives: In this study, the effects of the synchronic use of the stepwise profile dialysis fluid flow rate with increased blood flow rate (BFR) were studied on hemodialysis adequacy. The subjects received four routine hemodialysis sessions in group one and four hemodialysis sessions with the stepwise profile of the fluid flow rate with increased BFR in group two. Results: The mean score of dialysis adequacy was 0.89 in the routine method and 1.26 in the profile with increased BFR. Conclusion: The results of the study suggest that the stepwise profile of the dialysis fluid flow rate with increased BFR should be used synchronously to improve hemodialysis adequacy. According to the Iranian hemodialysis association, the number of patients with ESRD was 57 800 in 2016, of whom 29 200

Objectives
Methods
Results
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.