Abstract

There are several reasons for low intraoral curing light output as reported in the CRA Newsletter February 1989.l TRCINSFORNER STEPUP Output 126 VAC Fig. 1 illustrates the variation in light output versus line voltage input. The equipment used was a Cure Rite radiometer (EFOS, Williamsville, N. Y.), a Fluke 8921A True RMS voltmeter (John Fluke Mfg. Co., Everett, Wash.), a Visilux curing light (3M, St. Paul, Minn.) and a variable transformer. If low voltage is a problem, a wired small transformer will raise the alternating current line voltage; a 6-volt l-ampere transformer is commonly available (Fig. 2). If, after the wiring as shown is completed, the output is reduced by 6V (114 V), the leads marked x and y should be reversed. The output of the transformer needs to be rated for amperage rating of the curing light. X

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call