Abstract

Objectives: To compare radiation dose and image quality of three different CT devices using low dose abdominal protocols in patients with suspected urolithiasis. Differences in radiation dose due to usage of additional hardware (tin filter) and software (newer generation iterative reconstruction) were evaluated. Methods: Low - dose examinations from two standard CT (siemens AS +, siemens flash) devices with a tin - filter CT (siemens force) were compared regarding dose-length product (DLP) and computed tomography dose index volume (CTDIvol). Image quality of each CT scan was assessed using a five point Likert scale (0 = major blurring, 4 = excellent depiction). Kruskal - Wallis analysis was performed to test for significant subgroup differences in DLP, CTDIvol and image quality. An interrater agreement concerning image quality was evaluated. Results: CT examinations of a total of 143 patients were assessed. DLP of Force was 56% and 55% lower in comparison to AS+ and flash, (74.3, 166.1 and 164.6 mGycm respectively; P 0.05). Interrater agreement regarding image quality was substantial for all CT - devices (? = 0.75, n = 143). Conclusions: Modern CTs with the innovative technology of a built-in tin filter allow for significant reduction of radiation exposure in patients with suspected urolithiasis by optimizing the X - Ray spectrum with image quality as high as in standard CT devices

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.