Abstract

The authors analysed preoperative venogrphies of 3100 patients referred for transvenous lead extraction for a possible chance of safe haemodialysis catheter (HC) implantation or proper AVF function. A chance of safe catheter implantation parallel to existing leads reaches 68.8% ipsilaterally to CIED. Contraindications for implantation have been found in less than 2% of cases contralaterally. Ipsilaterally proper AVF function chance has been found in 50.3% of the cases and almost 98% contralaterally. A bilateral chest electrodes location require the special attention. Abandoned lead, lead burden, bilateral leads, additional lead implantation or abandonment, and implant duration may have a significant influence on HC insertion or proper function of arteriovenous fistula. (1) Obstruction of prominent thoracic veins is a frequent finding in CIED carriers and may impede or disable implantation haemodialysis accesses. (2) Implantation of temporary or permanent HC may be questionable ipsilaterally to the CIED in 31.2% and contralaterally in 2.0% of patients. Proper function of AVF is uncertain in 49.7% ipsilaterally and 2.1% contralaterally to CIED. (3) Pacing history and leads dwell time influence chances of success haemodialysis access even on the free-from CIED chest side. (4) Proper venous flow evaluation seems to be valuable in CIED carriers before an attempt of haemodialysis access formation, even contralaterally.