Abstract

To assess the reliability of the KidZ Health Castle formula (KHC-F) to determine the correct probe position of a multichannel intraluminal impedance pH. A retrospective cohort study was performed on 222 children between 1month and 18years of age undergoing multichannel intraluminal impedance pH. The primary outcome was the comparison of the pH sensor location determined by the KHC-F with the radiological target position. The margin of error was defined as 1 cm from the target position. Performance of the KHC-F and existing formulas was determined via the percentage with a correct position, mean error, 95% limits of agreement (Bland-Altman plots), and Spearman correlation. A post hoc analysis was performed with an updated KHC-F v2, subtracting -0.5cm from the KHC-F. Positioning with KHC-F was correct in two-thirds of the participants, with a very strong correlation (ρ = 0.91) with the target position. Bland-Altman plots showed good agreement between KHC-F and target position (mean error of -0.44cm, lower limit -3.2cm, upper limit 2.3cm). A post hoc analysis with the KHC-F v2 showed a correct positioning in 74% of patients. Comparison with other formulas showed a stronger performance of KHC-F and KHC-F v2 on correct positioning, mean error, and 95% limits of agreement. The KHC-F leads to reliable results. KHC-F v2 outperforms all other existing formulas in children, thereby reducing the need for repositioning and the amount of x-ray exposure. The age distribution of the sample may be a limitation, as well as the retrospective nature of the study.