Abstract
Thin antimony oxide covered AuSb alloy electrode was firstly found to be an excellent cold cathode for generating hot electrons during cathodic pulse polarization. Owing to the injection of hot electrons and the subsequent generation of hydrated electrons, fluorescein iso-thiocyanate (FITC) that cannot be excited in common ECL was cathodically excited at the alloy electrode. Self-assembled thiol monolayers were formed on the electrode surface due to the presence of Au in the alloy, to which strepavidin was covalently bound, and then biotinylated antibody was immobilized through the strepavidin–biotin interaction. As a simple model, an immunosensor for the detection of human IgG (hIgG) using FITC as labeling agent was fabricated. ECL signals were responsive to the amount of hIgG bounded to the immunosensor. The ECL intensity was linearly changed with the logarithm of hIgG concentration in the range of 1.0–1000 ng mL −1, and the detection limit was ca. 0.3 ng mL −1 (S/N = 3). The proposed immunosensor showed a broad linear range (three magnitudes), good reproducibility and stability, which is promising in detecting FITC-based labels in various types of bioaffinity assays.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.