Uncemented porous-coated cobalt—chrome femoral stems have produced satisfactory short-term clinical results, although slightly inferior to the results of cemented total hip arthroplasty. Proximal femoral bone resorption, osteolysis, and fibrous ingrowth have been reported with porous-coated stems. Hydroxyapatite-coated stems and titanium stems have been introduced to avoid these findings through improved fixation. Dual-energy x-ray absorptiometry allows quantitative examination of bone density changes after uncemented total hip arthroplasty. This study examines the host—bone response to hydroxyapatite-coated titanium and porous-coated cobalt—chrome stems of identical geometry in three paired bilateral cases at 5 to 7 years. Distinctly different patterns of femoral bone mineral density changes were observed with the two stem designs. The authors conclude that dual-energy x-ray absorptiometry is a promising technique allowing noninvasive analysis of uncemented stem fixation, and that the data from this study suggest improved fixation and stress transfer with hydroxyapatite-coated titanium stems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.