Abstract
The purpose of this study was to determine the hospital-wide incidence of bacterial contamination of point-of-care (bedside) ultrasound probes and coupling gel at a single academic medical center to predict the risk of nosocomial spread of infection. Bacterial cultures were performed on all point-of-care ultrasound probe surfaces and associated gel bottles in our institution (82 total probes in 9 separate departments). This process was repeated every 2 weeks for a total of 8 weeks; therefore, each probe was cultured 4 times during the study period. Of the 320 probe cultures, 18 (5.6%), resulted in positive growth, all of which identified nonpathogenic organisms common to human skin flora and the environment. No methicillin-resistant Staphylococcus aureus or other notable pathogens were identified. No gel cultures resulted in bacterial growth. Bacterial contamination of point-of-care ultrasound probes and coupling gel is low at this single academic medical center and involves nonpathogenic organisms only.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.