Abstract
We performed a method comparison between the Fujirebio® Lumipulse G AMH assay and the Roche® Elecsys AMH assay using the same pediatric samples. We described full pediatric gender and age-specific reference ranges for AMH using the Fujirebio® AMH assay on the Lumipulse G 600 II. The study was performed on 281 plasma samples collected in tubes with lithium heparin. The samples were from patients (135 males and 146 females) aged from 3 days to 22 years collected at the University Hospital Center of Tours. The Fujirebio® Lumipulse method showed excellent correlation with Roche® Elecsys but had a significant proportional positive bias. The data were used to propose pediatric reference values adapted to the Fujirebio® method. Our study described full pediatric gender and age-related reference ranges for AMH using the Fujirebio® AMH assay on the Lumipulse G600II. The delineation between normal male and female AMH concentrations make them valuable clinical tools for the monitoring of pediatric sexual and reproductive development from early childhood through the pubertal transition into adulthood.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.