Abstract

Dual-phase noble-gas time projection chambers (TPCs) have improved the sensitivities for dark matter direct search in past decades. The capability of TPCs to reconstruct 3-D vertexes of keV scale recoilings is one of the most advantageous features. In this work, we develop two horizontal position reconstruction algorithms for the PandaX-II dark matter search experiment using the dual-phase liquid xenon TPC. Both algorithms are optimized by the 83mKr calibration events and use photon distribution of ionization signals among photomultiplier tubes to infer the positions. According to the events coming from the gate electrode, the uncertainties in the horizontal positions are 3.4 mm (3.9 mm) in the analytical (simulation-based) algorithm for an ionization signal with several thousand photon electrons in the center of the TPC.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.