Abstract
Closed-loop insulin delivery is a promising option to improve glycaemic control and reduce the risk of hypoglycaemia. We aimed to assess whether overnight home use of automated closed-loop insulin delivery would improve glucose control. We did this open-label, multicentre, randomised controlled, crossover study between Dec 1, 2012, and Dec 23, 2014, recruiting patients from three centres in the UK. Patients aged 18 years or older with type 1 diabetes were randomly assigned to receive 4 weeks of overnight closed-loop insulin delivery (using a model-predictive control algorithm to direct insulin delivery), then 4 weeks of insulin pump therapy (in which participants used real-time display of continuous glucose monitoring independent of their pumps as control), or vice versa. Allocation to initial treatment group was by computer-generated permuted block randomisation. Each treatment period was separated by a 3-4 week washout period. The primary outcome was time spent in the target glucose range of 3·9-8·0 mmol/L between 0000 h and 0700 h. Analyses were by intention to treat. This trial is registered with ClinicalTrials.gov, number NCT01440140. We randomly assigned 25 participants to initial treatment in either the closed-loop group or the control group, patients were later crossed over into the other group; one patient from the closed-loop group withdrew consent after randomisation, and data for 24 patients were analysed. Closed loop was used over a median of 8·3 h (IQR 6·0-9·6) on 555 (86%) of 644 nights. The proportion of time when overnight glucose was in target range was significantly higher during the closed-loop period compared to during the control period (mean difference between groups 13·5%, 95% CI 7·3-19·7; p=0·0002). We noted no severe hypoglycaemic episodes during the control period compared with two episodes during the closed-loop period; these episodes were not related to closed-loop algorithm instructions. Unsupervised overnight closed-loop insulin delivery at home is feasible and could improve glucose control in adults with type 1 diabetes. Diabetes UK.
Highlights
We did this open-label, multicentre, randomised controlled, crossover study between Dec 1, 2012, and Dec 23, 2014, recruiting patients from three centres in the UK
Identical study insulin pump and real-time continuous glucose monitoring device were used during the two study periods, which were separated by a 3–4 week washout, during which participants used their own pump and discontinued continuous glucose monitoring
In all but three participants, closed loop improved time spent in target range. In one of those three participants, time spent in hypoglycaemia was reduced by 15·1% and in the other two by 2·4% and 2·7%
Summary
We did this open-label, multicentre, randomised controlled, crossover study between Dec 1, 2012, and Dec 23, 2014, recruiting patients from three centres in the UK. Patients aged 18 years or older with type 1 diabetes were randomly assigned to receive 4 weeks of overnight closed-loop insulin delivery (using a model-predictive control algorithm to direct insulin delivery), 4 weeks of insulin pump therapy (in which participants used real-time display of continuous glucose monitoring independent of their pumps as control), or vice versa. Allocation to initial treatment group was by computer-generated permuted block randomisation. Each treatment period was separated by a 3–4 week washout period. This trial is registered with ClinicalTrials.gov, number NCT0144014
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.