Abstract

Oculomotor disorders are known to have profound impacts on a patients' quality of life. However, current clinical practice lacks the capability to provide simultaneous assessment of three tightly coupled oculomotor control components, i.e. eye movement, lens accommodation, and pupil response. In this study, a holographic waveguide (HW) based benchtop optometer was constructed and evaluated with a model eye. Experimental result and quantitative analysis indicate that a HW can convey high quality retinal images to a camera at an illumination level safe for human subjects and support high accuracy measurements of ocular refractive error over a wide range. Further development of a HW-based system promises a wearable, see-through device for comprehensive assessment of oculomotor control components while the subject is engaged in normal daily activities and thus enable advanced research and clinical management of oculomotor disorders.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.