Abstract

Bisphenols and triclosan have been used in various products, and exposure to these chemicals may affect human health. The present study proposes a sensitive method for the determination of bisphenols A, F, S, and triclosan. The fiber was coated by amino/hydroxyl bifunctional microporous organic network and protected by polyvinylidene fluoride hollow fiber membrane for direct immersion solid phase microextraction. The limit of detection was 0.005 μg/L (μg/kg), and the recoveries were in the range of 76.7% to 107.5% (87.4% to 107.6%) for breast milk (infant formula), with intra-day and inter-day precisions <10.5% (7.3%) and 13.6% (8.4%), respectively. Fiber-to-fiber reproducibility of < 9.5% and a lifespan of >100 cycles were obtained. The 95th percentile estimated daily intake of total bisphenols was close to temporary tolerable daily intake for infants fed by human milk, which highlighted the needs for further attention on human exposure to BPA and its substitutes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call