Abstract

This document describes a microextraction technique using a porous polypropylene hollow fiber for the analysis of 27 emerging contaminants (ECs) in water samples by means of liquid chromatography-tandem mass spectrometry. The ECs under study were aniline, metformin, acetaminophen, carbazole, aldicarb, caffeine, thiabendazole, carbaryl, ibuprofen, atrazine, carbofuran, 2,4-dichlorophenoxyacetic acid, ametryn, bisphenol A, diuron, carbamazepine, methyl parathion, disulfoton, triclosan, diazinon, tebuconazole, malathion, chlorpyrifos, hydrocortisone, temephos, glibenclamide and 4 nitrophenol. The chromatographic separation was achieved in 13 min. The analyte detection was carried out using a triple quadrupole mass spectrometer, allowing the detection of a precursor ion and its fragmentation (product ions). The two-phase hollow fiber liquid phase microextraction conditions were optimized considering the parameters: organic solvent selection, sample pH, fiber length, sample volume, extraction time, stirring rate, sample temperature and salting-out effect. The developed method was validated. Detection and quantification limits were in the range of 1.09 to 98.15 ng L−1 and 2.13 to 126.50 ng L−1, respectively. The range of enrichment factor was from 6 to 4177. The method was successfully applied to water samples from various sources such as effluents from sewage treatment plants and surface water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.