Abstract

To obtain basic data about the holding power of a 4.5 mm AO/ASIF cortex screw in cortical bone in relation to bone mineral as expressed by densitometric methods, uniaxial pull-out tests were performed on 14 human cadaver femurs. The mechanical parameters were correlated with bone mineral which was assessed by quantitative computed tomography (QCT) and dual energy X-ray absorptiometry (DXA). High correlations were found between the QCT mass, the DXA density and content values and the holding power of the screw. The QCT density values, expressing the physical density of the bone, did not correlate similarly well.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.