Abstract

To establish the optimal flow cytometric method for HLA-B27 phenotyping, we compared several strategies, using three monoclonal anti-B27 antibodies (from the HLA-ABC-m3, GS145.2, and FD705 clones). We used a triple-color direct immunofluorescence assay, including a T-lymphocyte-specific antibody as an internal control and an anti-HLA-Bw4 antibody. Blood samples from >400 subjects were tested. From ROC curve analysis none of the three antibodies appeared to be suitable for use as a single typing reagent. The efficiency of the test was affected by cross-reactions with other HLA antigens, notably the HLA-B7 antigen. Preincubation with anti-B7 serum efficiently inhibited this cross-reaction and raised the test efficiency considerably. We concluded that none of the anti-B27 antibodies investigated is suitable for use as a single typing reagent. Additional typing of Bw4 is not valuable, whereas inhibition of cross-reactions due to the B7 antigen will considerably improve the performance of the test. We recommend that two different monoclonal anti-B27 antibodies be used for accurate and reliable HLA-B27 phenotyping with flow cytometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.