Abstract

The objective of this study was to evaluate the bone ingrowth of a new vacuum plasma sprayed titanium surface (vps-ti) in comparison to cs-titanium implants in a göttinger minipig model. Fifteen göttinger minipigs each received the two implants, vacuum plasma sprayed titanium with a porosity of 50% and a pore size of 200 microm (vps-ti) and an implant with a similar porosity but a different pore size 500 microm (cs-ti), at the proximal femur metaphysis by press-fit technique. The pigs were euthanized at three different postsurgical periods: 4, 8 and 12 weeks. Each femur was harvested and qualitative (macroscopic and microscopic) and quantitative (histomorphometric) histological analysis was done on histological slides. The results indicated that there was a difference in bone ingrowth between the two implants, whereas the bone ingrowth of vps-ti was superior to cs-ti after 4 and 8 weeks healing time. 12 weeks post implantationem no statistiscal difference was evident. The pore size of 200 microm seemed superior to a pore size of 500 microm. Whether or not these effects lead to a better mechanical stability remains unanswered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.