Abstract

Section thickness has been shown to affect the histomorphometrical measurement of bone-implant contact when analysed under brightfield microscopy. This study investigated whether microradiography of the bone-implant interface eliminated the errors associated with thick section analysis. Seven implant containing sections were utilized. Microradiographs of the thick (~100 μm) sections were taken and the sections were subsequently ground to thicknesses of 50 μm and 25 μm. Photomicrographs were taken of the microradiographs and of the sections at each thickness (100, 50 and 25 μm) under brightfield microscopy. The photomicrographs were analysed for direct bone-implant contact in the cortical passage region and along the total length of the implant. The effect of section thickness on multiple fluorochrome labelling in 10 rabbit femur specimens was also examined. Centre-to-centre interlabel distance was measured for each label pair at a thickness of 100 μm and then again after the sections were ground to 50 μm and 25 μm. The thick (100 μm) sections showed a significantly greater amount of bone-implant contact than either the thin sections or the microradiographs. There was no difference in direct bone-implant contact measured by microradiography or thin sections. However, the microradiographic analysis showed a much lower variability of the bone-implant contact than the sections evaluated under brightfield microscopy. In addition, they have the added benefit of providing information on bone mineral density. Centre-to-centre interlabel distance was not significantly different for any label pair owing to section thickness. Data from this study provides evidence that the use of microradiographs for histomorphometrical analysis of the bone-implant interface is superior to brightfield analysis of thin sections owing to the lower variability of microradiographical data and the ability to obtain bone mineral density measures. Additionally, given that interlabel distance was not significantly affected by section thickness, the use of 100 μm thick sections for analysis of fluorochrome labels in cortical bone is supported.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.