Abstract

Drill-free screws were developed to provide convenient orthodontic anchorage. The purpose of this study was to evaluate the effects of the drilling procedure on the stability of the screws under early orthodontic loading. Thirty-two screws were inserted into the jaws of 2 beagles. The screws were divided into 2 groups of 16: the drilling group and the drill-free group. Screws in the drilling group were inserted into the site that had been drilled with a pilot drilling bur, and those in the drill-free group were inserted without drilling. A force of 200 to 300 g was applied using nickel-titanium coil springs 1 week after insertion. Twelve weeks after insertion, mobility was tested with Periotest (Siemens AG, Bensheim, Germany), and the screws with the surrounding bone were prepared for histomorphometric evaluation. Screws in the drill-free group showed less mobility and more bone-to-metal contact; they had more bone area compared with the drilling group, although bone osseointegration was generally found in both groups. With careful technique, drill-free screws can provide stable orthodontic anchorage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.