Abstract

Matrix effects are considered to be a main obstacle of quantitative bioanalytical LC-MS/MS methods. Therefore it is often required to minimize them in order to increase method reliability. HILIC has been referenced as one of possible approaches. However, there is a lack of experimental evidence in scientific literature so far. Matrix effects were evaluated using spiked serum samples after SPE and protein precipitation prior to UHPLC-ESI-MS/MS. Chromatography was performed in both HILIC and reversed-phase mode. The influence of the matrix effects on the signal response was assessed using a set of 34 compounds of pharmaceutical interest and post-extraction addition approach. The advantages and drawbacks of the HILIC and reversed-phase chromatographic modes were compared and discussed in detail. HILIC demonstrated the potential to reduce the occurrence of matrix effects when a more thorough sample pretreatment procedure such as SPE was applied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.