Abstract
Interest in developing paper-based devices for point-of-care diagnostics in resource-limited settings has risen remarkably in recent decades. In this paper, we demonstrate what we refer to as "high yield passive rrythrocyte removal" (HYPER) technology, which utilizes capillary forces in a unique cross-flow filtration for the separation of whole blood with performance comparable to centrifuges. As we will demonstrate, state-of-the-art passive blood separation methods implemented in paper-based systems exhibit rapid blood cell clogging on the filtration media or serum outlet and yield only about 10-30% of the total serum present in the sample. Our innovation results from the inclusion of a differentiation pad, which exploits hydrodynamic effects to reduce the formation of a fouling layer on the blood filtration membrane resulting in more than 60% serum yield with undiluted whole blood as direct input. To demonstrate the effectiveness of the HYPER technology we implement it in a lateral flow system and demonstrate the accurate quantification of vitamin A and iron levels in whole blood samples in 15 minutes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.