Abstract

We set up an electronic gradiometer for magnetocardiography (MCG) in an unshielded environment. The electronically balanced gradiometer consists of high-temperature radio-frequency superconducting quantum interference device (rf SQUID) magnetometers. The rf SQUID magnetometers are arranged to form the electronic first-order, four-vector gradiometer, or second-order gradiometer. The output of the MCG signal was filtered by a band pass (1–30 Hz) and power line filters. We can reduce the noise to ∼1 pT Hz−1/2 at 1 Hz for second-order or four-vector gradiometers. Two-dimensional MCG imaging is demonstrated. The results are discussed with data measured in the moderate magnetically shielded environment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.