Abstract

Fundamental mechanical loss mechanisms exist even in very pure materials, for instance, due to the interactions of excited acoustic waves with thermal phonons. A reduction of these losses in a certain frequency range is desired in high precision instruments like gravitational wave detectors. Systematic analyses of the mechanical losses in those low loss materials are essential for this aim, performed in a highly sensitive experimental set-up. Our novel method of mechanical spectroscopy, cryogenic resonant acoustic spectroscopy of bulk materials (CRA spectroscopy), is well suited to systematically determine losses at the resonant frequencies of the samples of less than 10-9 in the wide temperature range from 5 to 300 K. A high precision set-up in a specially built cryostat allows contactless excitation and readout of the oscillations of the sample. The experimental set-up and measuring procedure are described. Limitations to our experiment due to external loss mechanisms are analysed. The influence of the suspension system as well as the sample preparation is explained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.