Abstract

Cardiac troponin (cTn) assays have quickly gained in analytical sensitivity to become what are termed 'high-sensitivity cardiac troponin' (hs-cTn) assays, bringing a flurry of dense yet incomplete literature data. The net result is that cTn assays are not yet standardized and there are still no consensus-built data on how to use and interpret cTn assay results. To address these issues, the authors take cues and clues from multiple disciplines to bring responses to frequently asked questions. In brief, the effective use of hs-cTn hinges on knowing: specific assay characteristics, particularly precision at the 99th percentile of a reference population; factors of variation at the 99th percentile value; and the high-individuality of hs-cTn assays, for which the notion of individual kinetics is more informative than straight reference to 'normal' values. The significance of patterns of change between two assay measurements has not yet been documented for every hs-cTn assay. Clinicians need to work hand-in-hand with medical biologists to better understand how to use hs-cTn assays in routine practice.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.