Abstract

This work describes a segmented radial turbo-spin-echo technique (DW-rTSE) for high-resolution multislice diffusion-weighted imaging and quantitative ADC mapping. Diffusion-weighted images with an in-plane resolution of 700 microm and almost free of bulk motion can be obtained in vivo without cardiac gating. However, eddy currents and pulsatile brain motion cause severe artifacts when strong diffusion weighting is applied. This work explains in detail the artifacts in projection reconstruction (PR) imaging arising from eddy currents and describes an effective eddy current compensation based on the adjustment of gradient timing. Application of the diffusion gradients in all three orthogonal directions is possible without degradation of the images due to eddy current artifacts, allowing studies of the diffusional anisotropy. Finally, a self-navigation approach is proposed to reduce residual nonrigid body motion artifacts. Five healthy volunteers were examined to show the feasibility of this method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.