Abstract

A commercial laboratory microwave acid digestion system was evaluated for the acid dissolution of ceramic powders (Al2O3, AlN, BN and Si3N4) prior to the determination of their trace element content by microwave induced plasma atomic emission spectrometry. Newly designed vessels, capable of withstanding internal pressures of over 110 bar, provide rapid and satisfactory results for sample dissolution. Sample preparation time was approximately 30 min (including the subsequent cooling time and preparation of the final solution). Results from conventional stainless-steel acid digestion vessel (“Teflon bomb”) dissolution are compared with the microwave bomb results of microwave plasma atomic emission spectrometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.