Abstract

In order to quantify 1,3-dinitropyrene (DNP), 1,6-DNP and 1,8-DNP in soil, we developed an efficient clean-up procedure and a sensitive determination method using fluorescence detection. DNP isomers were efficiently cleaned by three stages of fractionation, i.e., a silica gel open column chromatography using stepwise elution and two further purification steps by high-performance liquid chromatography (HPLC) using a monomeric-type octadecylsilyl (ODS) column and a polymeric-type ODS column. The recoveries of DNPs during the whole clean-up process were 94% or more. The fraction corresponding to DNPs was injected into an analytical polymeric-type ODS column for HPLC to separate DNP isomers. The effluent from the analytical ODS column was directly introduced to a catalyst column, which was packed with 5 μm alumina coated with platinum and rhodium (Pt–Rh), in order to reduce DNPs to diamino compounds, and then the fluorescence of diaminopyrenes was detected. The immediate detection of diaminopyrene isomers after on-line reduction afforded a sensitive detection of DNP isomers. The detection limits for DNPs were in the range of 0.7 to 4 pg. These developed methods were applied to four soil samples collected at parks in residential areas.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call