Abstract
A sensitive HPLC–electrochemical detection method was developed for the analysis of gliclazide (GL) in human plasma. After deproteination of 100 μL of plasma by acetonitrile, evaporation, and reconstitution, GL was separated on a C 18 column (150 mm × 4.6 mm) by the mobile phase (70 mM disodium tetraborate, pH 7.5, containing 26.5% of acetonitrile). The regression equations were linear ( r > 0.9990) over the range of 50 nM to 4.00 μM. The precision and accuracy of intra- and inter-day analysis were less than 5.3 and 0.93% for relative standard deviation and relative error, respectively. The limit of detection for plasma was 10 nM for GL (S/N = 3, 10 μL injection). This newly developed method was applied for monitoring blood levels with one healthy volunteer dosing with a GL tablet.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.