Abstract
The performance of a surface plasmon resonance (SPR) sensor with silver-gold bimetallic film based on angular interrogation was presented. To achieve minimal reflectance at resonance angle, the thicknesses of silver and gold layers of the bimetallic sensor were optimized at first. Then, the sensitivity and full-width-half-maximum (FWHM) of SPR sensors with optimum silver-gold thicknesses were investigated. With the increasing of the thickness of silver layer in bimetallic film, both the sensitivity and FWHM of the SPR sensor decreases, which indicates that the bimetallic sensor can obtain higher sensitivity than the SPR sensor with single silver layer, and smaller FWHM than the one with single gold layer. In addition, by analyzing the electric field inside the bimetallic SPR sensor, we studied the underlying physics of the improvement of the bimetallic sensor performance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.