Abstract

We report the first clinical application of a noncompetitive immunometric assay system that provides advantages for the rapid and robust assay of small molecules similar to those realized for larger molecules with two-site immunometric assays. This anti-immune complex assay is based on the interaction of a receptor such as a primary antibody with its ligand, such that new binding sites, recognizable by a secondary antibody, are formed. In this report the system is applied to the measurement of digoxin in serum. Utilizing an anti-complex antibody that recognizes a digoxin-bound primary antibody with affinity > 2000-fold over its binding to the primary antibody alone, we show that this anti-complex assay system provides a high-performance assay for serum samples, being conveniently simple (immobilized primary antibody binds digoxin and then labeled secondary antibody so that when excess unbound label is washed away the immunometric readout reflects the digoxin concentration), rapid (incubation time 1-10 min), sensitive (detection limit 30 ng/L), precise (3-4% within-run CV, 1-8% total CV), and free from interference from digoxin-like immunoreactive factors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.