Abstract

An adaptive optics scanning laser ophthalmoscope (AO-SLO) using a liquid-crystal spatial light modulator was developed. For routine clinical applications, long-term stability of the AO system is very important because unavoidable eye movement may degrade the instrument's performance. We studied the long-term performance of the aberration correction with healthy human eyes. Retinal image acquisition and AO data collection were performed simultaneously for periods of several minutes. We confirmed that, for more than 90% of the periods, the root-mean-square errors of residual wavefront were below the Marechal criterion. Drifts and microsaccades of fixational eye movement were examined using retinal images and residual aberrations. The results showed significant correlation between the transverse shift of retinal image and the low-order residual wavefront aberration during the drifts.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.