Abstract

BackgroundLong-term continuous systolic blood pressure (SBP) and heart rate (HR) monitors are of tremendous value to medical (cardiovascular, circulatory and cerebrovascular management), wellness (emotional and stress tracking) and fitness (performance monitoring) applications, but face several major impediments, such as poor wearability, lack of widely accepted robust SBP models and insufficient proofing of the generalization ability of calibrated models.MethodsThis paper proposes a wearable cuff-less electrocardiography (ECG) and photoplethysmogram (PPG)-based SBP and HR monitoring system and many efforts are made focusing on above challenges. Firstly, both ECG/PPG sensors are integrated into a single-arm band to provide a super wearability. A highly convenient but challenging single-lead configuration is proposed for weak single-arm-ECG acquisition, instead of placing the electrodes on the chest, or two wrists. Secondly, to identify heartbeats and estimate HR from the motion artifacts-sensitive weak arm-ECG, a machine learning-enabled framework is applied. Then ECG-PPG heartbeat pairs are determined for pulse transit time (PTT) measurement. Thirdly, a PTT&HR-SBP model is applied for SBP estimation, which is also compared with many PTT-SBP models to demonstrate the necessity to introduce HR information in model establishment. Fourthly, the fitted SBP models are further evaluated on the unseen data to illustrate the generalization ability. A customized hardware prototype was established and a dataset collected from ten volunteers was acquired to evaluate the proof-of-concept system.ResultsThe semi-customized prototype successfully acquired from the left upper arm the PPG signal, and the weak ECG signal, the amplitude of which is only around 10% of that of the chest-ECG. The HR estimation has a mean absolute error (MAE) and a root mean square error (RMSE) of only 0.21 and 1.20 beats per min, respectively. Through the comparative analysis, the PTT&HR-SBP models significantly outperform the PTT-SBP models. The testing performance is 1.63 ± 4.44, 3.68, 4.71 mmHg in terms of mean error ± standard deviation, MAE and RMSE, respectively, indicating a good generalization ability on the unseen fresh data.ConclusionsThe proposed proof-of-concept system is highly wearable, and its robustness is thoroughly evaluated on different modeling strategies and also the unseen data, which are expected to contribute to long-term pervasive hypertension, heart health and fitness management.

Highlights

  • Long-term continuous systolic blood pressure (SBP) and heart rate (HR) monitors are of tremendous value to medical, wellness and fitness applications, but face several major impediments, such as poor wearability, lack of widely accepted robust SBP models and insufficient proofing of the generalization ability of calibrated models

  • To measure the pulse transit time (PTT), i.e., the time delay for the pressure wave to propagate between two arterial sites, one popular method is based on two signals, i.e., electrocardiography (ECG) and photoplethysmogram (PPG)

  • We propose a single-arm blood pressure monitoring system, which allows for placing the PPG sensor and the ECG electrodes all on the left upper arm, to enable long-term daily applications which have critical requirements on the wearability and comfortableness

Read more

Summary

Introduction

Long-term continuous systolic blood pressure (SBP) and heart rate (HR) monitors are of tremendous value to medical (cardiovascular, circulatory and cerebrovascular management), wellness (emotional and stress tracking) and fitness (performance monitoring) applications, but face several major impediments, such as poor wearability, lack of widely accepted robust SBP models and insufficient proofing of the generalization ability of calibrated models. To measure the PTT, i.e., the time delay for the pressure wave to propagate between two arterial sites, one popular method is based on two signals, i.e., electrocardiography (ECG) and photoplethysmogram (PPG) The former one is the electrical signal generated by the heart, while the latter one measures fluctuations in the blood volume which are caused by the mechanical pressure pulse and changes later than the electrical ECG wave. The ECG signal is usually measured using a single- or multiple-lead configuration, referring to or modified from the traditional standard 12-lead configuration which can provide strong ECG signals with highly distinguishable morphologies [2, 3] These electrodes placement methods may have some limitations in long-term applications, e.g., the chest electrodes placement may be uncomfortable especially when sweating, and the two wrists configuration may be still inconvenient since additional wires or separate devices are inevitable. The PPG sensor is usually placed on the chest which may be uncomfortable, or on the finger where more challenges may be posed to the integration of PPG and ECG sensors [4]

Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.