Abstract
A detailed analysis of the effects of temperature changes, over time and between array elements, on the generation of circular artifacts in images produced by x-ray computed tomography was reported. We give formulas for calculating--according to the x-ray energy, detector sensitivity, and observed image quality (contrast and spatial resolution)--the maximum offset temperature coefficient and maximum gain temperature coefficient that will allow circular-artifact-free imaging. A temperature-controlled and insulated solid-state x-ray detector array, consisting of Gd2O2S:Pr,Ce,F ceramic scintillators coupled to crystal Si pin-photodiodes and designed to meet the requirements for these coefficients, produced high-resolution artifacts-free CT images of a phantom head.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.