Abstract

In order to investigate the contamination due to the radioactive isotopes (RIs) in NaI(Tl) scintillator for dark matter search, a simple method of pulse shape discrimination (PSD) to extract α-rays energy spectrum was developed. Two linear signals of different timings were input into a single module charge-sensitive analog-to-digital converter. The α- and the γ-rays were discriminated by comparing the total charge integration of a pulse with the partial one. As a result, the α-rays energy spectrum with NaI(Tl) was successfully and clearly extracted. The present work showed that the PSD method was performed by the minimum electronic circuits. The sensitivity was high enough to measure RIs of a few μBq/kg in high pure NaI crystals for dark matter search.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.