Abstract
This paper reports the use of a polymer inclusion membranes (PIMs) for direct determination of Al(III) ions in natural water by using a fluorescence based optode. The best composition of the PIMs consisted of 60 wt.% (m/m) poly (vinyl chloride) (PVC) as the base polymer, 20 wt.% (m/m) triton X-100 as an extractant, 20 wt.% (m/m) dioctyl phthalate (DOP) as plasticizer and morin as the reagent, was used in this study. The inclusion of triton X-100 was used for enhancing the sorption of Al(III) ions from liquid phase into the membrane phase, thus increasing the optode fluorescence intensity. The optimized optode was characterized by a linear calibration curve in the range from 7.41 × 10(-7) to 1.00 × 10(-4) molL(-1) of Al(III), with a detection limit of 5.19 × 10(-7) molL(-1). The response of the optode was 4 min and reproducible results were obtained for eight different membranes demonstrated good membrane stability. The optode was applied to the determination of Al(III) in natural water samples. The result obtained is comparable to atomic absorption spectrometry method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.