Abstract

This paper investigated a potential instrument for carbon disulfide in situ measurement with high precision and strong anti-interference capability. A compact and automated carbon disulfide detection system was developed using a fiber opto-electronic sensing device. A custom software interface based on LabVIEW was developed. The multi-wavelength least-squares method based on differential optical absorption spectroscopy was employed for improved detection and the anti-interference capabilities of the system. The detection limit of the system (signal-to-noise ratio=3) was determined to be 1ppb per meter. Using this scheme, the reliability of a liquid-seal was verified to have carbon disulfide leakage. Although the liquid level of the liquid-sealed carbon disulfide showed no significant change over 24h, a residue concentration of over 30ppm remained detectable on the surface.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.