Abstract

In this study, a convenient, sensitive, rapid and simple method was developed on microfluidic chip which was integrated with on-line complexing and laser-induced fluorescence detection. A rhodamine derivative (RD) was developed as a fluorescent chemosensor for Hg(II). It exhibited high selective recognition toward Hg(II) over other examined metal ions in water samples. Under the optimized conditions, the response was linearly proportional to the concentration of Hg(II) in the range of 0–70 μM with a detection limit of 0.031 μM. Satisfactory repeatability and reproducibility were achieved, with a relative standard deviation (RSD) of 6.62%. The established method was successfully applied for the determination of Hg(II) in environmental water samples (surface water, tap water, and waste water). Recoveries obtained for the determination of Hg(II) in spiking samples ranged from 85% to 103%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call