Abstract

A highly selective PVC membrane electrode based on a cerium-salen complex was prepared. The sensor displays an anti-Hofmeister selectivity sequence with a preference for iodide ion over many common organic and inorganic anions. The proposed electrode exhibits a near-Nernstian behavior over a wide concentration range (5.0 x 10(-2) - 8.0 x 10(-6) M) with a slope of 57.5 mV per decade, and a detection limit of 6.0 x 10(-6) M. The electrode has a very fast response time and can be used in the pH range of 3.0 - 1 1.0. It was applied, as an indicator electrode, in potentiometric titration of Ag+ ions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.