Abstract

A simple and selective high performance liquid chromatographic method has been developed for the determination of total iron in water samples. 2‐(8‐quinolylazo)‐4,5‐diphenylimidazole (QAI) was used as a precolumn chelating reagent for iron(II). The iron(II)‐QAI chelate was separated on a Cosmosil 5C8‐MS column with an aqueous acetonitrile mobile phase containing tetrabutylammonium bromide and spectrophotometrically detected at 710 nm. Iron(III) was determined as the iron(II) chelate with QAI after it was reduced to iron(II) using ascorbic acid. Detection limit, defined as three times of standard deviation of a blank signal, was 18 pg of iron(II) in 100 µL injection. Total iron in environmental water samples was selectively determined without any pretreatment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.