Abstract
Threshold values for 147 subjects (9-43 years old) were measured with a high-frequency (HF) air-conduction (AC) (Interacoustics AS 10 HF) and an electric bone-conduction (EBC) (Audimax 500) audiometer. In addition, the reproducibility of these methods was studied in another group of 24 subjects. The results confirmed the previous findings of Okstad et al. (1988) that the electric current (i) used as a stimulus in the Audimax 500 audiometer can be converted into decibels with a correction factor of 40 log (i) re 1 mA as Tonndorf & Kurman (1984) have proposed. However, an additive frequency-dependent correction is needed to obtain similar loudness sensation increases with these audiometers. Reproducibility with the EBC audiometer was better than with the AC audiometer, especially in the HF range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.