Abstract

An impedance read-out circuit with high accuracy implemented using a TSMC 0.18 um mixed-signal CMOS process for BIA-type (bio-electrical impedance analysis) sensors to detect biomarkers is proposed in this investigation. In addition to proposing the impedance estimation circuit, the associated analytic proof is also disclosed. Using a high-order approximation model and high noise rejection capability by the addition of OPAs (operational amplifiers), the proposed system demonstrates a solution with high accuracy in terms of the maximum impedance error ( $$\mathrm<$$ 6 %) and the least phase error ( $$\mathrm < $$ 3 $${^\circ }$$ ) in a frequency range from 1 kHz to 2 MHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.