Abstract

We have developed a novel, high-throughput approach for molecular mass determination of proteins from 14 200 to 116 000 based upon multiplexed, absorbance-based capillary electrophoresis. Via capillary multiplexing, 96 samples were analyzed simultaneously within 30 min. Detection with ultraviolet light obviates the need for protein staining or derivatization. The detection limit of the system was estimated at 5 μg/ml bovine serum albumin (BSA) when sampled from 12.5 mM Tris–HCl. The linear dynamic range was over two orders of magnitude from 5 μg/ml to 1000 μg/ml for BSA. Better than 5% sizing accuracy for protein molecular mass determination and excellent run-to-run and day-to-day reproducibility was obtainable with the described method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.