Abstract
A miniaturized immunoassay for human interleukin-13 (IL-13) using homogeneous time-resolved fluorescence (HTRF®) has been developed. In this assay, IL-13 which was secreted from NK3.3 cells stimulated with interleukin-2 (IL-2) was detected by measuring the time-resolved fluorescence after adding a mixture of three reagents, biotinylated anti-IL-13 monoclonal antibody, europium cryptate (fluorescence donor)-labeled different anti-IL-13 monoclonal antibody and crosslinked allophycocyanin (fluorescence acceptor)-conjugated with streptavidin in a 384-well assay plate. The detection limit of IL-13 using this immunoassay was estimated to be less than 600 pg/ml and IL-13 levels measured by this method were very close to those measured by enzyme linked immunosorbent assay (ELISA; the correlation coefficient was 0.9535). The proposed assay requires only a fourth of the quantities of all reagents compared with the assay using a conventional 96-well microtiter plate. Furthermore, there is no need to transfer the culture supernatant to another assay plate and wash the plate. Therefore, this miniaturized immunoassay is economical and efficient and is particularly suitable for high-throughput drug screening.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.