Abstract
Fluid flow in microchannels is used to treat or wash samples and can be incorporated into high-throughput applications such as drug screening, which currently use standard microtiter wells for performing assays. This paper provides theoretical and experimental data comparing microchannels and standard wells on the metrics of sample washing and experimental error in treatment concentrations. It is shown numerically and experimentally that microchannel concentration can be approximated with an inverse linear relationship to input volume. The experimentally supported mathematical approximation and error propagation methods are used to compare the accuracy and precision of treatments in microchannels vs. standard wells. Mathematical results suggest microchannels can provide 10 or more times the treatment precision of standard wells for volume ratios typical of high-throughput screening. Passive-pumping and diffusion are utilized to improve microchannel accuracy and precision even further in a treat-wait-treat method. The advantages of microchannels outlined here can have large-scale effects on cost and accuracy in screening applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.