Matrix assisted ionization vacuum (MAIV) rapidly generates gas-phase analyte ions from subliming solid-phase matrix:analyte crystals for analysis by mass spectrometry (MS). Ionization from the solid-phase allows the use of a variety of surfaces for introducing matrix:analyte samples to the vacuum of a mass spectrometer, including common laboratory materials, such as disposable pipet tips, filter paper, tooth picks, and nylon mesh. MAIV is shown here to be capable of analyses as fast as 3 s per sample with achievable sensitivities in the low femtomole range. MAIV-MS coupled with ion mobility spectrometry (IMS)-MS and tandem mass spectrometry (MS/MS) is shown to be especially powerful for analysis and characterization of a wide range of molecules ranging from small molecules such as drugs and metabolites (∼300 Da) to intact proteins (25.6 kDa). Automated sample introduction is demonstrated on two different commercial mass spectrometers using a programmable XYZ stage. A MAIV high-throughput nontargeted MS(E) approach is also demonstrated utilizing IMS for rapid characterization of small molecules and peptides from standard solutions, as well as drug spiked human urine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.